Posizione corrente:

Home / CT-100

Technical Specifications

Current Transformation Ratio - N 1:1000
Maximum DC Primary Current - IP(DC) ±100 A
Current Polarity Bipolar
Maximum DC Secondary Current - IS(DC) ±100 mA
Small Signal Bandwidth (±3 dB) - typ. BW > 500 kHz
> 200 kHz ("V"-version)
Equivalent Input noise (@Bandwidth) * < 1 ppm/FS @ 200 Hz
< 10 ppm/FS @ 50 kHz
Output Voltage ("V"-version) - VOUT ±10 V
Output Voltage Gain ("V"-version) - VOUT/IP(DC) 100 mV/A
Maximum Output Current ("V"-version) ±15 mA
Maximum external Shunt Resistor 0...40 Ω
Temperature Coefficient - TC < 0.5 ppm/°C
< 2 ppm/°C ("V"-version)
Maximum Total Output Offset* < 10 ppm / FS
Non-Linearity < 5 ppm
< 15 ppm ("V"-version)
Induction into Primary (typ.) 35 µV (RMS)
Protection Signal OK Status
Supply Voltage (±6%) ±15 V
Connections D-sub 9 connector ("C"-model)
7-pin strip TH type ("P"-model)
Mechanical (Outer) Dimensions 45 x 57 x 75 mm
Primary Conductor Hole Diameter - Ø 16 mm
Operating Temperature Range 0 ... 50 °C
Accuracy typ. * < 30 ppm / FS
< 0.25% / FS (“V” - version)
Maximum Weight 250 g
* These specifications are guaranteed only when the CT sensor is used with the dedicated CAEN ELS PS1215 power supply

Ordering Options

Product code Description Ordering Code
CT-100-C 100A Primary Current 0-FLUCS, DB-9 connector WCT100CXAAAA
CT-100-P 100A Primary Current 0-FLUCS, 7-pin PCB connector WCT100PXAAAA
CT-100V-C 100A Primary Current 0-FLUCS, DB-9 connector, Voltage Output WCT100VCXAAA
CT-100V-P 100A Primary Current 0-FLUCS, 7-pin PCB connector, Voltage Output WCT100VPXAAA
CT-0.05% CT - Voltage Output Accuracy < 0.05 % WCTCUST0500X

Overview
Overview
Immagine del prodotto

The 0-FLUCS (0-FLUx Current Sensor) family is based on a closed loop technology that allows accurate and precise monitoring of DC and AC currents with high bandwidth, very low noise and negligible drift.

The CT-100 transducers are rated at a at maximum bipolar primary current of 100A with a transformation ratio of 1:1000.

Galvanic isolation between the primary and the secondary circuits allows to measure currents at a different potential and simplifies interfacing when using the 0-FLUCS as the feedback element of current regulated power supplies.

Output from the CT-100 transducers can be chosen between two different versions: secondary current output or buffered voltage output (low temperature coefficient shunt resistor and low-noise amplifier are embedded in the device).

Also connection type can be chosen between the 'C' option – a male D-sub 9 Connector, also know as DE-9 or DB-9 – and the 'P' option – 7-pin through-hole for PCB mounting.

Main characteristics of the 0-FLUCS current transformers are negligible temperature coefficient on the secondary output current, excellent linearity and extremely low noise.

DC current transformers represents the ideal replacement for systems where Hall-effect sensors are used as current sensing elements and better performances are needed.

0-FLUCS exclusive technology allows the user to obtain performances which cannot be reached with other current sensing techniques.

Want to get more into this topic? Just click the following link:

0-FLUCS Exclusive Technology - An Overview


Immagine del prodotto

All CT-100 devices also have different mounting holes in order to be easily installed in different configurations. Both self-threading screws and normal ones can be used.

Main application fields for these current transducers are precise and extremely stable regulated power supplies and power inverters.

Due to the excellent characteristics, the CT-100 transducers can be used in a variety of calibration, acceptance testing and quality control applications in industrial, power generation and automotive fields.

Commercially available versions of the CT-150 current transformers are the current-output CT-100-C and CT-100-P with their respective voltage- output versions CT-100V-C and CT-100V-P. An LED indicating the correct operation of the device is present on all versions.


Features
  • Monitoring of AC and DC currents
  • Excellent Linearity
  • Closed-loop detection
  • Galvanically isolated from primary
  • Low Temperature Drift
  • Current-output or Voltage-output versions
  • Wide bandwidth
  • High accuracy
  • UL 94 V-0 flammability grade
  • LED indicates correct operation
  • D-sub 9 connector or 7-pin PCB-mount versions
  • 3-year extended warranty

Applications
Documentation Software/Firmware
CT-100-150 - User's Manual v1.3 Last update: 2016-07-15 13:42:15

CT-100-150 - Product Presentation Last update: 2017-03-21 13:00:16

0-FLUCS Technology - Introduction v1.2 Last update: 2015-07-01 12:17:38